Fluke 36 manual
Digital multimeter. Service Manual. Comments 1. Topic: Fluke Page 1 of 1. Add Comment. Notify me of new comments via email. I have read and agree to the terms and conditions. By submitting a comment, you are declaring that you agree with these rules: Although the administrator will attempt to moderate comments, it is impossible for every comment to have been moderated at any given time.
Referring to Table 4, put the Meter in the function and range for Function Test number 1 i. Apply the first indicated output i.
The reading on the Meter display should be within the low and high values shown i. Repeat steps for each function and range in Table 4. If the Meter fails to perform within the low-high range indicated for each test in Table 4, the Meter needs to be calibrated and adjusted, or requires some repair.
Welcome to ManualMachine. We have sent a verification link to to complete your registration. Log In Sign Up. Forgot password? Enter your email address and check your inbox.
Please check your email for further instructions. Enter a new password. Fluke 36 Service Manual Download for 1. Use caution when working with voltages above 60V dc or 30V ac. W Warning To avoid false readings, which could lead to possible electric shock or personal injury, replace the battery as soon as the battery indicator appears.
Y Caution The Meter is protected throughout by double insulation or reinforced insulation. Caution This Meter contains parts that can be damaged by static discharge. PN September Rev. To order replacement parts, call Outside U. Service, repairs, and replacement parts are warranted for 90 days.
General Specifications General specifications are shown in Table 1. Table 1. General Specifications. Required Equipment Table 1. General Specifications cont. Functional Specifications Functional specifications are shown in Table 2. Table 2. Table 3. Required Equipment Equipment Recommended Model.
Performance Test P Warning To avoid shock, do not execute the performance tests unless the Meter is fully assembled. Table 4. Press MAX button.